Request Quote | Specification Sheet | FAQ | Reference List

Notes for MP406 Soil Moisture Sensor & Interfaces


Part No.   Description
SMD4-P Interface for MP406 Soil Moisture Sensor (VSW%) Output
Range of Response:  0 to 100 VSW% (air to water)
VSW% (Volumetric Soil Water Content)
Output Display: 0 to 100 VSW%
Display is 000.0 VSW% default (1 decimal place) and 000.00 VSW% in high resolution mode (2 decimal places).
Note: The data is always recorded as 000.00 (high resolution) in the Smart Logger.
Output Calibration: VSW% calibration using the lookup table as supplied in the MP406 manual and below for agricultural soils which are typically saturated at 50-55 VSW%.
Accuracy:
±2 VSW% 0-55 VSW% i.e. 50 ±2%
±10 VSW%     55-100 VSW%     i.e.     80 ±10%
 
SMD4-V Interface For MP406 Soil Moisture Sensor (mV) Output
Range of Response: 0 to 100% (air to water)
Output Display: 0 to 1,200 mV (millivolt)
Display is 000.0 mV default (1 decimal place) and 000.0 mV in high resolution mode (1 decimal place).

VSW% mV MP406 VSW% mV MP406
0.00   0       55.00   1015      
2.00   120       60.00   1025      
5.00   210       65.00   1035      
10.00   310       70.00   1045      
15.00   415       75.00   1055      
20.00   510       80.00   1065      
25.00   610       85.00   1075      
30.00   720       90.00   1085      
35.00   825       95.00   1105      
40.00   895       100.00   1160      
45.00   955          
50.00   1005          

  • Reading Rate. The switching of the MP406 connected to the SMD4-P can be 10, 20, 30 or 60 minutes for the economy mode of operation. These are set when the logging frequency is set in the datalogger. This will suit the typical field user. A laboratory column user can, every few minutes, log water flow in columns if the SMD4-P is set to non economy mode. This necessitates the use of a power supply to the DataBus from a wall charger (part No. CH7A).

  • Cycle time of the SMD4-P Interface depends on the soil moisture content. If all MP406 are in wet soil, the cycle time is 5 seconds per channel or 20 seconds total. If all of the MP406 are in dry soil, the cycle time is 2 seconds per channel or 8 seconds total. Warm up time is 0.5 seconds for each MP406.

    Note: the cycle times apply whether 1, 2, 3 or 4 MP406 are connected to the SMD4-P

  • Unregulated power supply to the MP406 must exceed 7 volts and be less than 15 volts. This means that an MP406 system will always require a 12 volt battery to be connected to the Databus.

  • Calibration is done using a precision voltage calibration standard. The SMD4-P has the MP406 interface calibrated with the Look Up Table data as set out in the MP406 manual. Calibration and the SMD4-V will be simply the millivolt output.

  • Staggered starting times of 1 minute delay is supplied with consecutively numbered interfaces supplied with a system (1 minute / serial no. advance).
Related Documents
PDF SMD Soil Moisture Sensor Interface Instruction Manual 810K