Request Quote | Specification Sheet | FAQ | Reference List

GT3-1 Custom Tensiometer
Transducers



Measuring Principle

The soil water tension is transmitted via the ceramic cup into the tensiometer. The vacuum inside the tensiometer is measured by a pressure transducer, which gives a continuous analogue output signal.

Resolution and Accuracy

A resolution of 0.1 kPa (1 cm H2O) can be attained for the GT3 Tensiometer Transducer. The accuracy actually obtained is related to the calibration and resolution of the AD converter.

In field applications the values obtained should consider the shaft/cup length of the tensiometer: 1 cm = 0.1 kPa.


Transducer connection

Red +10 V power supply
Yellow Signal output "+"
Pressure change from 0 to -7 kPa
Output from 0 to -20 mV
Black Battery ground
Blue Signal output "-"

Ordering Information

GT3-1 Transducer with 1/4 inch NPT thread, 1 m cable
ST1 Optional interface for use with ICT Smart Logger

Note: Customers requiring cable lengths greater than 1 m are requested to source cable locally. Standard 4 wire cable is required.

 